TAM30 Series
30 Watts Medical GradeAC/DC Power Adapters
TRUMPower
FEATURES
• 90~264 VAC Universal Input
• Class I with IEC 320/C14 AC Inlet
• Power-on LED Indicator
• Low Earth Leakage Current
• Standby Consumption < 0.3W
• Meet CEC & Energy Star Efficiency Level V Requirements
MODELS LIST
Average Efficiency (Typical)
Output
Voltage
Min. Output
Current
Max. Output
Current
Ripple & Noise
(mVp-p)*
Max. Output
Power
Product No.
Tolerance
@ 115 VAC
83%
@ 230 VAC
81%
TAM30-05
TAM30-09
TAM30-12
TAM30-15
TAM30-18
TAM30-24
TAM30-48
5V
9V
0A
0A
0A
0A
0A
0A
0A
5.0A
3.33A
2.5A
5%
5%
5%
5%
5%
5%
5%
75mV
90mV
25W
30W
30W
30W
30W
30W
30W
86%
84%
12V
15V
18V
24V
48V
120mV
150mV
180mV
240mV
480mV
86%
85%
2.0A
88%
86%
1.66A
1.25A
0.625A
88%
86%
89%
87%
91%
90%
* Measured at rated line voltage and output load ranges, 20 MHz BW, with a 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor across the output.
INPUT SPECIFICATIONS
OUTPUT SPECIFICATIONS
Output Power Ratings ......... See table
Ripple and Noise ................ See table
Overvoltage Protection ........ Set at 116-230% of its nominal
output voltage
Input Voltage Range ............ 90~264 VAC
Input Frequency ................... 47~63Hz
Input Current ........................ 1.0A (rms) @ 115 VAC
0.6A (rms) @ 230 VAC
Overcurrent Protection ........ Protected to short-circuit conditions
Transient Response........... Max. excursion of 4% or better on
all models, recovering to 1% of
Inrush Current ..................... 50A @ 115 VAC or 100A @ 230 VAC,
at 25 C cold start
Earth Leakage Current ........ 100µA max. @ 264 VAC, 63Hz
°
final value within 500 µs after a
25% step load change
GENERALSPECIFICATIONS
Efficiency ............................. See table
STANDARDS & COMPLIANCES
Hold-up Time....................... 8 ms min. at 115 VAC
Turn on Delay Time ............. 3 seconds max. at 115 VAC
Switching Frequency ........... 67 KHz
IEC 60601-1-2 ..................... EMC performance
EN 55011, FCC, VCCI ......... Class B, conducted & radiated
EN 61000-3-2...................... Harmonic distortion, Class A
EN 61000-3-3...................... Line flicker
EN 61000-4-2...................... ESD, 8 KV air and 6 KV contact
EN 61000-4-3...................... Radiated immunity, 3V/m
EN 61000-4-4...................... Fast transient/burst, 2 KV
EN 61000-4-5...................... Surge, 1 KV diff., 2 KV com.
EN 61000-4-6...................... Conducted immunity, 3 Vrms
EN 61000-4-8...................... Magnetic field immunity, 3A/m
EN 61000-4-11 .................... Voltage dip immunity,
30% reduction for 500ms,
Line Regulation ................... 0.5% max. at full load
°
°
Operating Temperature ...... 0 C to +50 C
°
Derating............................... Derate from 100% at +40 C linearly
°
to 50% at +60 C
°
°
Storage Temperature .......... -40 C to +85 C
Relative Humidity ................ 10% to 90% non-condensing
Withstand Voltage ............... 4,000 VAC from input to output
MTBF .................................... 300K hrs min. at full load,
°
25 C ambient, calculated per
MIL-HDBK-217F
60% reduction for 100ms,
>95% reduction for 10ms
Safety Standards ................. UL60601-1, EN 60601-1
CSA C22.2 No. 601.1
Agency Approvals ................ UL, cUL, TUV, CB, CE (LVD)
Other Compliances ............ RoHS, CEC & Energy Star Level V
Tumbler Technologies + TRUMPower
3350 Scott Blvd., Building 13 • Santa Clara, California 95054, USA • 408-988-6616 • sales@trumpower.com • www.TRUMPower.com