Catalog: 1654001
Issue Date: 06.2011
Corcom Product Guide
Power Inlet Line Filter for Medical Equipment
H Series
UL Recognized
CSA Certified
VDE Approved*
EH8
EH9
EH4
EH1
EH5
H Series
Specifications
• Minimal leakage current suitable for medical
equipment
Maximum leakage current each Line to Ground:
@ 120 VAC 60 Hz:
2 A
5 A
@250 VAC 50 Hz:
• Two element circuit provides basic EMI
attenuation above 1 MHz
Hipot rating (one minute):
Line to Ground:
Line to Line:
2250 VDC
1450 VDC
• Available with an internal ground circuit inductor
(C suffix versions) to isolate equipment chassis
from power line ground at radio frequencies
3
Rated Voltage (max.):
250 VAC
50/60 Hz
3 to 15A*
• Flanged mounting the same as the EC, ED and EF
Series
Operating Frequency:
Rated Current:
• Capacitive output (see EAH, EBH and EJH Series
for capacitive input)
Operating Ambient Temperature Range
(at rated current I ):
-10°C to +40°C
In an ambient temperature (T ) higher than +40°C
r
Ordering Information
3 E H 1 C
a
the maximum operating current (I ) is calculated as
o
follows: I = I (85-Ta)/45
o
r
Ground Circuit Inductor
Omit if none
Electrical Schematic
LINE
LOAD
Output Styles
1
Chassis Mount Style
1 – .250 [6.3] spade terminals
3 – Wire leads
Power Inlet Style *
4 – .250 [6.3] spade terminals
Available Part Numbers
(back)
5 – Wire leads (bottom)
3EH1
3EH3
6EH1
6EH3
6EH4
6EH5
6EH8
8 – Wire leads (back)
9 – .250 [6.3] spade terminals
(bottom)
6EH9
10EH1
10EH3
10EH4
15EH4
H Series
Ground Circuit Inductor Versions
10EH4C
Leakage current designation
E – Low Leakage (<0.5 mA)
Current Rating
3, 6, 10 or 15A
*15A versions are tested by Underwriters Laboratories
to US and Canadian requirements and are
VDE approved at 10A, 250VAC
*IEC 60320-1 C14 inlet mates with C13 connector
169
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.